SPRINT FIDELIS
Report
- Report Number
- 2649622-2010-13911
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S029
- Removal / Correction Number
- Z-0067-0070-2008
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. ANALYSIS OF THE LEADS IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE SPRINT FIDELIS LEAD MODEL IS INCLUDED IN A FIELD ADVISORY. EVALUATION SUMMARY: (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. EVALUATION SUMMARY: (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION CONDUCTOR WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON ALL CONDUCTORS (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE OUTER TUBING OVERLAY WAS MELTED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IMPEDANCE HAS DROPPED FROM 1000 OHMS TO 500 OHMS SINCE IMPLANT 5 YEARS AGO. THE LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IMPEDANCE HAS DROPPED FROM 1000 OHMS TO 500 OHMS SINCE IMPLANT 5 YEARS AGO. IT WAS FURTHER REPORTED THAT THERE WAS HIGH THRESHOLD ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE PATIENT WAS IN CHRONIC ATRIAL FIBRILLATION, AND THAT DURING EXTRACTION OF THE RV LEAD, THE ATRIAL LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6949 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Hospitalization| O| R | 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| A7700 HEART VALVE| A7700 HEART VALVE| A7700 HEART VALVE| 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD| 7288 IMPLANTABLE PACEMAKER/CARDIO/DEFIB |