ENRHYTHM DR
Report
- Report Number
- 6000144-2010-06322
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS REVEALED LOW TELEMETERED BATTERY VOLTAGE. ON (B)(6) 2010, THE TELEMETERED BATTERY VOLTAGE WAS 2.61 VOLTS (V) WHILE THE DAILY BATTERY VOLTAGE MEASUREMENT WAS 2.93 V. (B)(4) THE ACTUAL DEVICE WAS ANALYZED - CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS REVEALED LOW TELEMETERED BATTERY VOLTAGE. ON (B)(6) 2010, THE TELEMETERED BATTERY VOLTAGE WAS 2.61 VOLTS (V) WHILE THE DAILY BATTERY VOLTAGE MEASUREMENT WAS 2.93 V.
IT WAS REPORTED THAT THERE HAVE BEEN VARYING BATTERY VOLTAGE MEASUREMENTS. IT WAS LATER REPORTED THAT THERE WAS A OBSERVATION OF IMMINENT ELECTIVE REPLACEMENT INDICATOR (ERI) DURING DEVICE INTERROGATION AND THERE WAS POSSIBLE PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED WITH A NEW DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THERE HAVE BEEN VARYING BATTERY VOLTAGE MEASUREMENTS. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| O| R | 5024M IMPLANTABLE PACING LEAD| 4269 COMPETITOR IMPLANTABLE PACING LEAD| 4269 COMPETITOR IMPLANTABLE PACING LEAD| 5024M IMPLANTABLE PACING LEAD |