FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1925469 · Received December 14, 2010

Report

Report Number
2649622-2010-13882
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
October 4, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): NO ANOMALIES FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED. BLOOD WAS IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD DURING IMPLANT HAD AN IMPEDANCE OF 2500OHMS. THERE WAS AN ATTEMPT TO REPOSITION THE LEAD BUT THE HELIX APPEARED TO NOT BE FULLY RETRACTED. THERE WAS AN ATTEMPT TO RE-EXTEND THE HELIX BUT IT WOULD NOT DEPLOY. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 86 YR Other