Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THE DEFIBRILLATION COIL WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE OUTER TUBING OVERLAY MELTED AND HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VENTRICULAR LEAD WAS REMOVED DUE TO HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.