VIRTUOSO DR
Report
- Report Number
- 6000144-2010-06296
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S062
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS ANALYZED AND NO ANOMALIES WERE FOUND. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND INDICATED HIGH RESISTANCE/IMPEDANCE. AN OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE MEASUREMENT INITIATED A PATIENT ALERT (B)(6) 2010. RV PACE LEAD IMPEDANCE RISES FROM AN APPROXIMATE BASELINE OF 375 OHMS THE WEEK OF (B)(6) 2010 TO A HIGH OF 1904 OHMS THE WEEK OF (B)(6) 2010.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THERE WAS AN INCREASE IN IMPEDANCE ON THE RIGHT VENTRICULAR LEAD (COMPETITOR). THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THERE WAS AN INCREASE IN IMPEDANCE AND A HIGH/UNSTABLE/UNMEASUREABLE THRESHOLD ON THE RIGHT VENTRICULAR LEAD (COMPETITOR). THE DEVICE WAS EXTRACTED ALONG WITH THE LEAD. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VIRTUOSO DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D154AWG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD |