FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 1925360 · Received December 14, 2010

Report

Report Number
6000144-2010-06296
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S062
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS ANALYZED AND NO ANOMALIES WERE FOUND. PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED AND INDICATED HIGH RESISTANCE/IMPEDANCE. AN OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE MEASUREMENT INITIATED A PATIENT ALERT (B)(6) 2010. RV PACE LEAD IMPEDANCE RISES FROM AN APPROXIMATE BASELINE OF 375 OHMS THE WEEK OF (B)(6) 2010 TO A HIGH OF 1904 OHMS THE WEEK OF (B)(6) 2010.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN INCREASE IN IMPEDANCE ON THE RIGHT VENTRICULAR LEAD (COMPETITOR). THE DEVICE AND LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS AN INCREASE IN IMPEDANCE AND A HIGH/UNSTABLE/UNMEASUREABLE THRESHOLD ON THE RIGHT VENTRICULAR LEAD (COMPETITOR). THE DEVICE WAS EXTRACTED ALONG WITH THE LEAD. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D154AWG ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD