FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1925191
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13744
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S003
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR WAS DISTORTED AND HAD BLOOD/BODY FLUID, THE OUT INSULATION COSMETIC ESC AND DEPRESSION. APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THE LEAD HAD HIGH OUTPUT, HIGH THRESHOLD AND INTERMITTENT CAPTURE. THE PHYSICIAN ATTEMPTED TO REPOSITION THE LEAD BUT WAS UNABLE TO DO SO. THE LEAD WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4193 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4076 IMPLANTABLE PACING LEAD |