ENDOPATH** XCEL
Report
- Report Number
- 3005075853-2010-07084
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- November 17, 2010
- Report Date
- November 17, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GCJ
- PMA / PMN Number
- K032676
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). DUCKBILL, LEAK TEST FAILURE THE ANALYSIS RESULTS FOUND THAT DEVICE (A) WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE ANALYSIS RESULTS FOUND THAT DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. THE FINAL QUALITY RELEASE CRITERIA WERE MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION.
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION WAS REQUESTED.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTRECTOMY PROCEDURE, AN AIR LEAK OCCURRED WITH A "BOO" SOUND WHEN INSTRUMENTS WERE REMOVED THROUGH THE DEVICES DURING USE. THE AIR LEAK STOPPED WHEN INSTRUMENTS WERE INSERTED THROUGH THE DEVICES. SINCE THE AIR DID NOT LEAK TOO MUCH, THE DEVICES WERE USED AS-IS TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH** XCEL | LAPAROSCOPE, GENERAL AND PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |