FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 1924809 · Received December 10, 2010

Report

Report Number
1811755-2010-01993
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
May 21, 2010
Report Date
May 21, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS EVALUATED BY A FIELD SERVICE TECH AND THE EVENT WAS CONFIRMED. THE WIRES WITHIN THE CONNECTOR WERE WORN OFF. THE WIRES WERE REPAIRED AND THE DEVICE WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELECTRICAL PLUG SPARKS WHEN PLUGGING AND UNPLUGGING UNIT. IT IS UNK IF THERE WAS PT INVOLVEMENT OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO FYD STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK