FDA Adverse Event
Malfunction
Summary report: N
NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE
MDR report key: 1924809
·
Received December 10, 2010
Report
- Report Number
- 1811755-2010-01993
- Event Type
- Malfunction
- Date Received
- December 10, 2010
- Date of Event
- May 21, 2010
- Report Date
- May 21, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- FYD
- PMA / PMN Number
- K012991
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS EVALUATED BY A FIELD SERVICE TECH AND THE EVENT WAS CONFIRMED. THE WIRES WITHIN THE CONNECTOR WERE WORN OFF. THE WIRES WERE REPAIRED AND THE DEVICE WAS RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ELECTRICAL PLUG SPARKS WHEN PLUGGING AND UNPLUGGING UNIT. IT IS UNK IF THERE WAS PT INVOLVEMENT OR ADVERSE CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE | AIR-HANDLING APPARATUS FOR A SURGICAL OPERATING RO | FYD | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |