FDA Adverse Event Malfunction Summary report: N

ZEVO CERVICAL PLATE SYSTEM

MDR report key: 19246988 · Received May 6, 2024

Report

Report Number
1030489-2024-00399
Event Type
Malfunction
Date Received
May 6, 2024
Date of Event
February 7, 2024
Report Date
May 6, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HXX
UDI-DI
00643169342293
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

E: FIRST NAME, LAST NAME OF INITIAL REPORTER AND FACILITY DETAILS ARE UNKNOWN. H3: PRODUCT ANALYSIS OF PART# 3030007, LOT# IM15C046 VISUAL AND OPTICAL INSPECTION REVEALED THE TIP OF THE DRIVER IS BROKEN OFF. DAMAGE PRESENT O THE TIP IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A USER FACILITY (UF) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT IS USED IN UNKNOWN SPINAL THERAPY. IT WAS REPORTED THAT THE DRIVER IS STRIPPED AND IT WILL NO LONGER HOLD A SCREW FOR IMPLANTATION.. THERE IS NO PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764204 ZEVO CERVICAL PLATE SYSTEM SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC 3030007 IM15C046 00643169342293

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown