FDA Adverse Event
Malfunction
Summary report: N
MICRO OSCILLATING SAW
MDR report key: 1924689
·
Received December 9, 2010
Report
- Report Number
- 1811755-2010-01989
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 22, 2010
- Report Date
- November 22, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- GFA
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS A MALFUNCTIONING 3RD PARTY MOTOR AND PROBLEMS WITH THE PRESS PLUG. THOSE PARTS WERE REPLACED ALONG WITH OTHER COMPONENTS. SERVICE DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING A SURGICAL PROCEDURE. THE CASE WAS COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO OSCILLATING SAW | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | GFA | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |