FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1924661 · Received December 10, 2010

Report

Report Number
2531779-2010-02800
Event Type
Malfunction
Date Received
December 10, 2010
Date of Event
November 9, 2010
Report Date
November 9, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS WITHOUT THE COMPLAINANT BATTERY. THE BATTERY COMPARTMENT AND CAP WERE FOUND TO BE INTACT WITH NO EVIDENCE OF MOISTURE INGRESS. EVAL FOUND THAT THE PUMP POWERS UP PROPERLY WITH A TEST BATTERY. THE PUMP WAS EXERCISED FOR 3 HOURS WITH NO EVIDENCE OF OVERHEATING OR ALARMS. HIGH CURRENT DRAWS WERE FOUND DURING TESTING, A CAPACITOR IN THE PIEZO CIRCUIT WAS FOUND TO HAVE FAILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FAMILY MEMBER NOTICED THAT THE BATTERY WAS HOT WHEN SHE REMOVED IT FROM THE PUMP DURING A ROUTINE BATTERY CHANGE. SHE STATED THAT IT WAS NOT HOT ENOUGH TO BURN THE PT. THE FAMILY MEMBER SAID THAT SHE INSPECTED THE BATTERY COMPARTMENT, SHE DENIED MOISTURE OR CORROSION IN COMPARTMENT, AND SHE STATED THAT THE BATTERY DID LOOK CORRODED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 12 YR