CORE IMPACTION DRILL
MDR report key: 1924657
·
Received December 9, 2010
Report
- Report Number
- 1811755-2010-01979
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 19, 2010
- Report Date
- November 19, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MFR FOR EVAL, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH BEARINGS, WHICH WERE REPLACED ALONG WITH OTHER COMPONENTS. SERVICE DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING DURING AN EXTRACTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER DEVICE. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.