FDA Adverse Event
Malfunction
Summary report: N
MICRO SAGITTAL SAW
MDR report key: 1924654
·
Received December 9, 2010
Report
- Report Number
- 1811755-2010-01991
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 23, 2010
- Report Date
- November 23, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MFR FOR SERVICE, AND A CONDITION OF RUN-ON WAS FOUND. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS PROBLEMS WITH THE MOTOR, PRESS PLUGS, SPHERICAL BEARING, AND BLADE MOUNT. THOSE PARTS WERE REPLACED ALONG WITH THE HOUSING. SERVICE REPAIRED AND RETURNED THE DEVICE TO THE CUSTOMER.
Description of Event or Problem · 1
THE HANDPIECE WAS RETURNED TO THE MFR FOR SERVICE, AND DURING THE INVESTIGATION A RUN-ON CONDITION WAS FOUND. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE. THERE WAS NO PT INVOLVEMENT OR ANY ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO SAGITTAL SAW | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |