CAPSUREFIX NOVUS
MDR report key: 1924626
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13653
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD BIPOLAR IMPEDANCE WAS LOW, AND IT WAS ALSO LOWER THAN THE LEAD'S UNIPOLAR IMPEDANCE, POSSIBLY DUE TO INNER INSULATION BREAKDOWN. THE LEAD IS STILL IN USE, AND THE PHYSICIAN MAY REPROGRAM THE LEAD PACING AND SENSING CONFIGURATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Required Intervention | (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |