FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1924626 · Received December 14, 2010

Report

Report Number
2649622-2010-13653
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD BIPOLAR IMPEDANCE WAS LOW, AND IT WAS ALSO LOWER THAN THE LEAD'S UNIPOLAR IMPEDANCE, POSSIBLY DUE TO INNER INSULATION BREAKDOWN. THE LEAD IS STILL IN USE, AND THE PHYSICIAN MAY REPROGRAM THE LEAD PACING AND SENSING CONFIGURATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Required Intervention (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD