FDA Adverse Event
Injury
Summary report: N
CAPSURE SENSE
MDR report key: 1924568
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13620
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- August 28, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P830061/S034
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND; FULL LEAD WAS RETURNED AND ANALYZED. BLOOD/BODY FLUID ON PROXIMAL CONDUCTOR(NOT OBSTRUCTED). OUTER INSULATION BREACHED CUT. APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT ONE WEEK AFTER IMPLANT THE RIGHT VENTRICULAR LEAD LOST CAPTURE AND HAD AN OUT OF RANGE HIGH IMPEDANCE. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SENSE | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4074 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |