FDA Adverse Event
Injury
Summary report: N
ATTAIN OTW
MDR report key: 1924543
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13606
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR PACING POLARITY WAS REPROGRAMMED AND THE LEFT VENTRICULAR LEAD IMPEDANCE DECREASED FROM 600 TO 300 OHMS. IT WAS ALSO REPORTED THAT THE DEVICE DISPLAYED NOISE ON THE ATRIAL AND VENTRICULAR CHANNELS, POSSIBLY DUE TO ELECTRO-MAGNETIC INTERFERENCE. BOTH THE LEAD AND DEVICE ARE STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other| R | 6940 IMPLANTABLE PACING LEAD| 6942 IMPLANTABLE TACHY LEAD |