FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 1924526
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13598
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- September 30, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): HELIX DISENGAGED FROM HELICAL CHANNEL, AND BLOOD NOTED ON HELIX MECHANISM; THE ANALYST NOTED THAT THE HELIX HAD BEEN OVER-RETRACTED; FULL LEAD RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT ATTEMPT THE LEAD HELIX WOULD NOT EXTEND ("MOVE OUT"). THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |