Malfunction
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1924496
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13585
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- September 28, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S017
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY; (B)(4): THE FULL LEAD WAS RETURNED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD ON/IN THE HELIX MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT OF THE RIGHT VENTRICULAR LEAD, THE HELIX EXTENSION WAS INCONSISTENT. R-WAVE AMPLITUDE IN THE PATIENT WOULD DECAY FROM 8-10 MV TO 2-3 MV AT BOTH THE APEX AND SEPTAL WALL LOCATIONS. IMPEDANCE AND THRESHOLDS WERE GOOD. A DIFFERENT LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.