FDA Adverse Event
Injury
Summary report: N
SPRINT
MDR report key: 1924440
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13564
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- September 22, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S13
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD'S SENSE VALUE SUDDENLY DECREASED, AND THAT THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK DUE TO T-WAVE OVERSENSING. THE LEAD WAS ABANDONED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6943 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | Hospitalization| L| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |