FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 1924440 · Received December 14, 2010

Report

Report Number
2649622-2010-13564
Event Type
Injury
Date Received
December 14, 2010
Date of Event
September 22, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD'S SENSE VALUE SUDDENLY DECREASED, AND THAT THE PATIENT RECEIVED AN INAPPROPRIATE SHOCK DUE TO T-WAVE OVERSENSING. THE LEAD WAS ABANDONED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6943 ASKU

Patients

Seq Age Sex Outcome Treatment
1 38 YR Hospitalization| L| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB