FDA Adverse Event Malfunction Summary report: N

NONE

MDR report key: 19243909 · Received May 3, 2024

Report

Report Number
2955842-2024-14195
Event Type
Malfunction
Date Received
May 3, 2024
Date of Event
May 20, 2023
Report Date
April 4, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
10886874111045
PMA / PMN Number
K112263
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE A DA VINCI PRODUCT TO PERFORM FAILURE ANALYSIS. THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY WAS ANALYZED AND FOUND TO HAVE THERMAL DAMAGE. IT EXHIBITED LOCALIZED MELTING WITH A CRATER-LIKE HOLE APPROXIMATELY 0.12" X 0.10" AT THE SILICON OVERMOLD, WHICH IS INDICATIVE OF ARCING. THE MCS INSTRUMENT REQUIRES THE USE OF A TIP COVER TO PREVENT ENERGY FROM DISCHARGING ANYWHERE BUT AT THE INSTRUMENT¿S TIP.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) FOLLOWED UP WITH THE INITIAL REPORTER AND OBTAINED ADDITIONAL INFORMATION. THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY THAT WAS RETURNED WAS USED IN THE SAME PROCEDURE AND WAS RELATED TO THE SAME EVENT.

Description of Event or Problem · 0

THE MONOPOLAR CURVED SCISSORS (MCS) TIP COVER ACCESSORY WAS RETURNED WITH NO REPORTED ISSUE.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
758553 NONE TIP COVER ACCESSORY NAY INTUITIVE SURGICAL, INC 400180-14 L80220930 10886874111045

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose DA VINCI INSTRUMENTS AND ACCESSORIES