SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2010-13510
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- September 27, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- Z-0474-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THE SHAFT SEAL WAS OUT OF POSITION. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD) AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. THERE WERE NO ANOMALIES WERE FOUND. THE SHAFT SEAL WAS OUT OF POSITION. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD) AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD DISLODGED AFTER ITS IMPLANT THE PREVIOUS DAY. THE LEAD WAS REPOSITIONED, BUT DEFIBRILLATION TESTING WAS UNSUCCESSFUL AT 14 J AND 25 J; IT WAS SUCCESSFUL AT 35 J. THE LEAD WAS REPOSITIONED AGAIN AND DEFIBRILLATION TESTING WAS UNSUCCESSFUL AGAIN AT 14 J AND 25 J. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR | Hospitalization| L| R | D224VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB| D224VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB| D224VRC IMPLANTABLE PACEMAKER/CARDIO/DEFIB |