FDA Adverse Event
Malfunction
Summary report: N
UNIVERSAL DRILL
MDR report key: 1924208
·
Received December 9, 2010
Report
- Report Number
- 1811755-2010-01946
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- November 18, 2010
- Report Date
- November 18, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RECEIVED BY THE MFR AND THE COMPLAINT WAS CONFIRMED. INTERNAL COMPONENTS WERE EVALUATED AND EXTENSIVE CORROSION WAS DISCOVERED WITHIN THE MOTOR ASSEMBLY AND BEARINGS. THE PRESENCE OF CORROSION CAN LEAD TO INCREASED FRICTION BETWEEN ROTATING COMPONENTS, RESULTING IN HEAT BEING GENERATED. THE MOTOR ASSEMBLY AND BEARINGS WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED BY THE MFR SALES REP, THE DRILL HEATED UP. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNIVERSAL DRILL | DRILLS, SURGICAL, ENT (ELECTRIC OR PNEUMATIC), INC | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |