CAPSURE SP NOVUS
MDR report key: 1924138
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13421
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P850089/S38
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED). THE OUTER INSULATION HAD A WHITE SUBSTANCE ON IT AS WELL AS A COSMETIC DEPRESSION. ONLY VISUAL ANALYSIS WAS PERFORMED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD UNACCEPTABLE THRESHOLD AND HIGH IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE SP NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5092 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| R | 5594 IMPLANTABLE PACING LEAD| KDR701 IMPLANTABLE PULSE GENERATOR |