FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 1924118
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13404
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- September 23, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCE AND INTERMITTENT CAPTURE. IT WAS ALSO REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL WITH SYNCOPE AND THE PATIENT HAD A FALL. THERE WERE PAUSES SEEN IN THE RHYTHM AND PATIENT HAD VT/VF. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |