FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1924096 · Received December 14, 2010

Report

Report Number
2649622-2010-13391
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD BEEN LOST TO FOLLOWUP, AND THAT THE DEVICE COULD NOT BE INTERROGATED AND WAS AT END OF LIFE WHEN THE PATIENT CAME IN FOR A DEVICE CHANGEOUT. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS REPORTED THAT THE LEAD HAD HIGH IMPEDANCE, NOISE, OVERSENSING, NO SENSING, NO CAPTURE, AND AN APPARENT FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| R 5054 IMPLANTABLE PACING LEAD