FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 1924085 · Received December 14, 2010

Report

Report Number
2649622-2010-13383
Event Type
Injury
Date Received
December 14, 2010
Date of Event
September 24, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S10
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD FRACTURED. THE SHORT INTERVAL COUNT (SIC) WAS ELEVATED AND THE LEAD INTEGRITY ALERT (LIA) WAS TRIPPED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6932 ASKU

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R 6940 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB