FDA Adverse Event Injury Summary report: N

ENPULSE

MDR report key: 1924083 · Received December 14, 2010

Report

Report Number
6000094-2010-02241
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S030
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND.

Description of Event or Problem · 1

ASKU

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD BEEN EXPERIENCING DIZZINESS AND SHORTNESS OF BREATH, WHICH RECURRED OVER AND OVER, AND THAT THERE WERE INTERMITTENT VENTRICULAR CAPTURE FAILURES AND UNSTABLE LEAD IMPEDANCE TRENDS. THE LEAD WAS TESTED AND FOUND TO HAVE NO ISSUE. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENPULSE IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. E2VDD01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R