FDA Adverse Event
Malfunction
Summary report: N
ATTAIN OTW
MDR report key: 1924061
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-13364
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- March 24, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P010015/S012
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) NO ANOMALIES FOUND; FULL LEAD ANALYZED. BLOOD/BODY FLUID PRESENT ON ALL CONDUCTORS. (B)(4) NO ANOMALIES FOUND; FULL LEAD ANALYZED. LEAD STRETCHED; LEAD DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATTAIN OTW LEAD WAS CONTAMINATED AND THE ATTAIN ABILITY LEAD WAS UNABLE TO BE POSITIONED. BOTH LEADS WERE NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN OTW | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4194 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Other |