SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2010-13359
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- September 21, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY, THE OUTER TUBING OVERLAY BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM, THERE WAS A TIP SEAL OBSERVATION AND THERE WAS APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD PERFORATED ONE DAY POST-OP. NO SERIOUS PATIENT INJURIES WERE OBSERVED BUT THE PATIENT COMPLAINED OF SHARP CHEST PAIN. THE LEAD WAS REPLACED WITH A NEW RV LEAD. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization| L| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |