FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1923927
·
Received December 13, 2010
Report
- Report Number
- 2531779-2010-02819
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 6, 2010
- Report Date
- November 11, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS. EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALED LOSS OF PRIME ALARMS ASSOCIATED WITH ZERO FORCE. REWIND, LOAD, AND PRIME STEPS WERE PERFORMED SUCCESSFULLY AND THE PUMP WAS EXERCISED WITH NO ALARMS OCCURRING.
Description of Event or Problem · 1
EVAL REVEALED A MISALIGNED DISPLAY SCREEN AND DISLODGED FORCE SENSOR PINS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |