FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 3000 SYSTEM
MDR report key: 1923915
·
Received December 9, 2010
Report
- Report Number
- 3002158293-2010-01310
- Event Type
- Malfunction
- Date Received
- December 9, 2010
- Date of Event
- September 21, 2010
- Report Date
- December 6, 2010
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF BATTERY PACK (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY WON'T HOLD CHARGE) WAS CONFIRMED. THE CAUSE OF THE BATTERY NOT FULLY CHARGING WAS A BROKEN WHITE WIRE ON THE BATTERY PCA BOARD WHERE THE WIRE ATTACHES TO THE BATTERY CELLS. THE ROOT CAUSE OF THE BROKEN WIRE HAS NOT BEEN DETERMINED, BUT MAY HAVE BEEN CAUSED BY A SEVERE SHOCK FROM DROPPING THE PACK. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACK. THE PT RECEIVED A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A PT SERVICE REP (PSR) CONTACTED ZOLL LIFECOR CUSTOMER SUPPORT TO REPORT THAT THE BATTERY OF A (B)(6) MALE PT WON'T HOLD A CHARGE. THE PT WAS PROVIDED WITH A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 3000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 3000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |