SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA
Report
- Report Number
- 2023826-2010-01228
- Event Type
- Malfunction
- Date Received
- December 1, 2010
- Date of Event
- October 26, 2010
- Report Date
- November 3, 2010
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- P900048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): EVAL CONCLUSION (OTHER) - AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MFG OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. BASED ON THE COMPLAINT HISTORY, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).
THE REPORTER STATED, THE SURGEON NOTED THAT A PIECE OF THE AQ2015A THREE PIECE SILICONE LENS WAS MISSING AFTER IT WAS INSERTED INTO THE EYE. THE LENS WAS REMOVED AND ANOTHER SAME MODEL LENS WAS IMPLANTED WITHOUT ANY PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | AQ2015A | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | INJECTOR: MODEL MSI-TM, LOT# UNK| CARTRIDGE: MODEL AQ CARTRIDGE-FP, LOT# UNK |