FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 1923775 · Received December 1, 2010

Report

Report Number
2023826-2010-01212
Event Type
Malfunction
Date Received
December 1, 2010
Date of Event
November 2, 2010
Report Date
November 3, 2010
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4) (NO LENS MALFUNCTION). EVAL, METHOD: LENS WORK ORDER SEARCH. RESULTS: VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND THE LENS OPTIC AND ONE HAPTIC TORN INTO TWO PIECES. THE OTHER HAPTIC WAS TORN OFF AND MISSING. THE LENS WAS RETURNED IN LIQUID. A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. CONCLUSIONS: BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVAL OF THE RETURNED PRODUCT, THE MOST LIKELY CAUSE OF THE EVENT WAS PT RELATED. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON INSERTED A (B)(4) COLLAMER ASPHERIC SINGLE PIECE LENS BUT THE CAPSULE WOULD NOT SUPPORT THE LENS. THE LENS WAS CUT IN HALF TO REMOVE THE PT'S EYE, WITH NO PT INJURY, NO ENLARGED INCISION OR SUTURES. THE REPORTER INDICATED THAT THE CAUSE OF THE EVENT MAY POSSIBLY BE PT RELATED, NOT LENS RELATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. CC4204A NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR INJECTOR: MODEL AND LOT NUMBER UNK| CARTRIDGE: MODEL AND LOT NUMBER UNK