ETS FLEX ARTICNG
Report
- Report Number
- 3005075853-2010-07050
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- November 10, 2010
- Report Date
- November 15, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(6). PINION AXLE SUPPORT. THE ANALYSIS RESULTS FOUND THAT THE ATS45 DEVICE WAS RETURNED WITH THE FIRING MECHANISM DAMAGED AND WITH NO RELOAD PRESENT. NO FUNCTIONAL TEST COULD BE PERFORMED DUE TO THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE CONDITION OF THE INTERNAL COMPONENTS AND THE FIRING TRIGGER TEETH AND THE LEFT SHROUD PINION AXLE SUPPORT WERE FOUND DAMAGED. WHILE NO CONCLUSION COULD BE REACH AS TO WHAT CAUSED THE FIRING MECHANISM TO FAIL, IT IS POSSIBLE THAT THE DEVICE WAS FIRED ON TISSUE THICKER THAN INDICATED OR THROUGH A LOCKED CARTRIDGE. WHEN EITHER OR BOTH OF THESE SCENARIOS OCCUR, THE COMPONENTS IN THE FIRING MECHANISM CAN BE DAMAGED. IT SHOULD BE NOTED THAT A 100% INSPECTION TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4). NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, AS THERE WAS NO RELOAD RETURNED FOR ANALYSIS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE (B)(4) DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE WAS RECEIVED FOR ANALYSIS. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A CYSTECTOMY PROCEDURE, THE DEVICE LOCKED OUT ON THE SECOND FIRING. THEY USED SUTURE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ETS FLEX ARTICNG | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |