PRISM REACTION TRAYS
Report
- Report Number
- 1415939-2010-00576
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Report Date
- November 18, 2010
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- JJE
- PMA / PMN Number
- K970042
- Removal / Correction Number
- 1415939-11/23/10-005-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRISM 6 CHANNEL ANALYZER, LIST # 6A36-04, SERIAL # (B)(4). THE CAUSE OF THE PRISM ERROR CODES FOR THE CALIBRATORS, CONTROLS OR SAMPLES WAS DUE TO AN INCREASE IN DRAIN TIME ERRORS AFFECTING SEVERAL LOTS OF PRISM REACTION TRAYS. A PRODUCT CORRECTION LETTER WAS SENT TO ABBOTT CUSTOMERS WITH INSTRUCTIONS TO DISCONTINUE USE OF THE PRISM REACTION TRAYS WHEN USED IN CONJUNCTION WITH THE PRISM HIV O PLUS OR HEPATITIS B SURFACE ANTIGEN ASSAYS. THE CUSTOMER LETTER ALSO INDICATED THE USE OF PRISM REACTION TRAYS WAS ACCEPTABLE IF NOT UTILIZING THE HIV O PLUS OR HEPATITIS B SURFACE ANTIGEN ASSAYS AND TO CONTINUE USE OF THE AFFECTED REACTION TRAYS UNTIL A NEW LOT OF REACTION TRAYS ARRIVED AT THE CUSTOMER FACILITY.
THE CUSTOMER OBSERVED 60 OCCURRENCES OF PRISM DRAIN TIME ERRORS IN A SINGLE RUN WHEN PRISM REACTION TRAY LOT 90359M500 WAS IN USE. THE CUSTOMER OBSERVED THE ERROR CODE ON BOTH ANALYZERS IN THE LAB. THERE WAS NO IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRISM REACTION TRAYS | REACTION TRAYS FOR PRISM CHEMILUMINESCENT IMMUNOASSAYS | JJE | ABBOTT LABORATORIES | 90359M500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PRISM 6 CHANNEL ANALYZER, LIST # 6A36-04 |