FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 1923597 · Received December 10, 2010

Report

Report Number
1222780-2010-00193
Event Type
Injury
Date Received
December 10, 2010
Date of Event
November 11, 2010
Report Date
November 11, 2010
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SERIAL NUMBER OF THE RADIO FREQUENCY CONTROLLER NOT PROVIDED BY THE COMPLAINANT. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT AND SERIAL NUMBERS. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY, UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. BASED ON THE INFO OBTAINED TO DATE, NO DIRECT CORRELATION CAN BE MADE BETWEEN THE REPORTED EVENT AND THE NOVASURE SYSTEM. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

DURING AN ATTEMPTED NOVASURE ENDOMETRIAL ABLATION PROCEDURE, THE PHYSICIAN HAD DIFFICULTY SEATING THE DEVICE. THE PROCEDURE WAS ABANDONED AND A PERFORATION IN THE FUNDUS WAS CONFIRMED ON POST HYSTEROSCOPY. NO TREATMENT WAS NEEDED FOR THE PERFORATION AND THE PATIENT WAS DISCHARGED AT HOME. ON (B)(6) 2010, IT WAS REPORTED THAT THE PATIENT HAS BEEN SEEN ON F/U AND WAS "FINE". A HYSTEROSCOPY WAS PERFORMED PRIOR TO THE ATTEMPTED ABLATION, AS WAS A DILATATION AND SOUNDING WITH A METAL SOUND (NOT A HOLOGIC DEVICE). IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 10E18RB

Patients

Seq Age Sex Outcome Treatment
1 49 YR Other RADIO FREQUENCY CONTROLLER - SERIAL NUMBER UNKNOWN