FDA Adverse Event Injury Summary report: N

EASYTOUCH

MDR report key: 19234742 · Received May 2, 2024

Report

Report Number
3005798905-2024-03181
Event Type
Injury
Date Received
May 2, 2024
Date of Event
May 2, 2024
Report Date
May 2, 2024
Manufacturer
FEEL TECH BIO CO., LTD.
Product Code
FMF
PMA / PMN Number
K102178
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CMO WAS NOT ABLE TO LINK DEVICE MALFUNCTION TO MANUFACTURING, THUS, EFFECTIVENESS CHECK IS NOT APPLICABLE. PREVENTATIVE MEASURES, HOWEVER, WERE TAKEN - PER CAPA-24-2024 DUE TO SIMILARITIES IN COMPLAINTS FROM CAPA-24-2024 AND CAPA-26-2024 - I.E., ADJUSTING AIR PRESSURE IN NEEDLE SILICONE APPLICATION PROCESS AND INCREASED FREQUENCY OF AIR PRESSURE OPERATION STATUS INSPECTION, AND RETRAINING OF STAFF.

Additional Manufacturer Narrative · 0

INITIAL TREND ANALYSIS FOR LOT 60204A WAS CONDUCTED, NO MALFUNCTIONS WERE FOUND. THIS IS THE ONLY COMPLAINT FOR LOT 60204A. FURTHER INVESTIGATION WILL BE CONDUCTED TO DETERMINE THE ROOT CAUSE OF COMPLAINT.

Additional Manufacturer Narrative · 0

CMO WAS NOT ABLE TO LINK DEVICE MALFUNCTION TO MANUFACTURING, THUS, EFFECTIVENESS CHECK IS NOT APPLICABLE. PREVENTATIVE MEASURES, HOWEVER, WERE TAKEN - PER CAPA-24-2024 DUE TO SIMILARITIES IN COMPLAINTS FROM CAPA-24-2024 AND CAPA-26-2024 - I.E., ADJUSTING AIR PRESSURE IN NEEDLE SILICONE APPLICATION PROCESS AND INCREASED FREQUENCY OF AIR PRESSURE OPERATION STATUS INSPECTION, AND RETRAINING OF STAFF.

Description of Event or Problem · 0

END USER REPORTED THAT WHILE REACHING INSIDE OF A BOX OF SYRINGES 830165 LOT 60204A THEIR HAND WAS PUNCTURED THROUGH THE POLYBAG. USER DID NOT SEEK MEDICAL ATTENTION DUE TO THE EVENT.

Description of Event or Problem · 0

END USER REPORTED THAT WHILE REACHING INSIDE OF A BOX OF SYRINGES 830165 LOT 60204A THEIR HAND WAS PUNCTURED THROUGH THE POLYBAG. USER DID NOT SEEK MEDICAL ATTENTION DUE TO THE EVENT.

Description of Event or Problem · 0

END USER REPORTED THAT WHILE REACHING INSIDE OF A BOX OF SYRINGES 830165 LOT 60204A THEIR HAND WAS PUNCTURED THROUGH THE POLYBAG. USER DID NOT SEEK MEDICAL ATTENTION DUE TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
847077 EASYTOUCH SYRINGE FMF FEEL TECH BIO CO., LTD. 60204A

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other