FDA Adverse Event Injury Summary report: N

OT ULTRA METER

MDR report key: 1923462 · Received December 13, 2010

Report

Report Number
2939301-2010-10688
Event Type
Injury
Date Received
December 13, 2010
Date of Event
November 22, 2010
Report Date
December 1, 2010
Manufacturer
LIFESCAN INC.
Product Code
NBW
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP # 2/SUPPLEMENTAL REPORT TEXT- 1/24/2011: THE LAY USER/PATIENT¿S TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED IN THIS CASE HAVE PASSED ALL TESTING WITH NO FAULTS FOUND. THE RETAIN TEST STRIPS ALSO PASSED ALL TESTING. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW UP # 1/SUPPLEMENTAL REPORT TEXT 12/20/2010. THE LAY USER/PATIENT'S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER INVOLVED IN THIS CASE HAS PASSED TESTING WITH NO FAULTS FOUND. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A SECOND FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT. 510(K) # IS K062195.

Description of Event or Problem · 1

ON (B)(6) 2010, A REPORTER FOR THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING THE PATIENT'S ONETOUCH ULTRA METER READ INACCURATELY HIGH COMPARED TO ANOTHER DEVICE. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN 3 WEEKS AGO PRIOR TO CONTACTING LFS. ON THE MORNING OF (B)(6) 2010, THE REPORTER STATED THE PATIENT OBTAINED A BLOOD GLUCOSE RESULT OF "367 MG/DL" WITH THE SUBJECT METER AND "45 MG/DL" ON ANOTHER METER, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <= 30% AND/OR <= 30 MG/DL. THE CCA WAS ADVISED THE PATIENT MANAGES HIS DIABETES WITH INSULIN (TYPE/ AMOUNT NOT SPECIFIED). THE REPORTER STATED THE PATIENT TOOK AN INCREASED DOSE OF INSULIN (10-12 UNITS). THE REPORTER WAS UNABLE TO SPECIFY ANY SYMPTOMS THE PATIENT DEVELOPED. THAT SAME MORNING, EMERGENCY MEDICAL SERVICES (EMS) WAS CONTACTED AND ADMINISTERED THE PATIENT INTRAVENOUS (IV) GLUCOSE AS TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT AND AN APPROVED SAMPLE SITE WAS USED FOR TESTING. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO TEST THE SUBJECT METER. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT ALLEGEDLY RECEIVED MEDICAL INTERVENTION FROM A HEALTH CARE PROFESSIONAL (HCP) AFTER THE REPORTED ISSUE BEGAN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE MISFIRED. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO OTHER DETAILS OF THE EVENT ARE KNOWN. THERE WAS NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT ULTRA METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN INC. 2961622

Patients

Seq Age Sex Outcome Treatment
1 49 YR Life Threatening| R