COLLEAGUE SINGLE CHANNEL VOLUMETRIC PUMP CE FRENCH
Report
- Report Number
- 6000001-2010-05779
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- July 6, 2010
- Report Date
- July 26, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DEPLETED BATTERY SET ALARM WAS DISCOVERED AND CONFIRMED BY BAXTER PERSONNEL IN THE PUMP'S EVENT HISTORY. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES. THIS CONDITION WAS RESOLVED AT THE FACILITY BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.06.00.
DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET ALARM, WHICH INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.