CELL-DYN SAPPHIRE ANALYZER
Report
- Report Number
- 2919069-2010-00580
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Report Date
- November 22, 2010
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISION
- Product Code
- GKZ
- PMA / PMN Number
- K051215
- Removal / Correction Number
- 2919069-4/7/10-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). EVALUATION: ASPIRATION BOTTOM SENSOR WORN-OUT FROM NORMAL USE. AN INVESTIGATION WAS CONDUCTED TO EVALUATE THIS ISSUE. NO PRODUCT DEFICIENCY WAS IDENTIFIED AND THE ISSUE WAS RELATED TO A WORN-OUT PART DUE TO NORMAL USE. A MANDATORY TECHNICAL SERVICE BULLETIN (TSB) IS ISSUED WITH INSTRUCTIONS FOR FIELD SERVICE TO INSPECT AND REPLACE THE PART WHEN NECESSARY. THE PREVENTIVE MAINTENANCE PROCEDURES WILL BE UPDATED IN THE OPERATORS MANUAL TO INCLUDE INSPECTION OF THE ASPIRATION PROBE SENSOR EVERY SIX MONTHS AND REQUEST REPLACEMENT IF REQUIRED. A DESIGN IMPROVEMENT IS IN PROCESS TO IMPROVE THE CABLE AND SEAL THE BOARD AND SWITCH ASSEMBLIES. A CORRECTION THROUGH A FOLLOW-UP CORRECTION AND REMOVAL (B)(4) 2010 WAS ISSUED TO INCLUDE INSTALLING A NEW SOFTWARE (V4). THE NEW V4 SOFTWARE RELEASED WITH (B)(4) 2010 INCLUDES A DESIGN IMPROVEMENT FOR THE ASPIRATION BOTTOM SENSOR. WITH V4, THE CD SAPPHIRE WILL EXECUTE SENSOR CHECKS TO DETERMINE THE STATUS OF THE SENSOR AND WILL GENERATE AN ERROR MESSAGE IF AN ISSUE IS DETECTED.
THE CUSTOMER REPORTED AN ISSUE WITH POOR POSITIONING OF THE PIPETTOR AND INSUFFICIENT SAMPLE VOLUME. TROUBLE SHOOTING REVEALED A VERTICAL DISPLACEMENT OF THE PROBE DESPITE RUBRICATING THE ASPIRATION ASSEMBLY. THE ISSUE WAS RESOLVED BY REPLACING THE CELL-DYN SAPPHIRE ASPIRATION BOTTOM SENSOR. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN SAPPHIRE ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKZ | ABBOTT DIAGNOSTICS DIVISION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ASPIRATION BOTTOM SENSOR |