FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER
MDR report key: 1923302
·
Received December 13, 2010
Report
- Report Number
- 3005075853-2010-07046
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- November 15, 2010
- Report Date
- November 18, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED SIX CONFORMING CLIPS AND THREE SCISSOR CLIPS. HOWEVER, THE JAWS WERE NOTED TO BE IN GOOD CONDITION. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE FINDING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE TIP OF THE CLIPS CROSSED OVER ON THE INITIAL FIRING. THERE WAS NO TORQUE APPLIED. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP** ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | G4TP4M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |