CAPSUREFIX NOVUS
MDR report key: 1923243
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-13321
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DISTAL PORTION OF THE LEAD WAS RETURNED AND ANALYZED. THE LEAD WAS RETURNED STRETCHED AND THE INNER INSULATION WAS BUCKLED WHICH WOULD PREVENT THE HELIX FROM FUNCTIONING PROPERLY. THE OUTER INSULATION WAS BREACHED CUT, AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM; APPARENT EXPLANT DAMAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD HELIX WOULD NOT EXTEND. THE LEAD WAS NOT IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |