FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1923141 · Received December 13, 2010

Report

Report Number
2649622-2010-13265
Event Type
Malfunction
Date Received
December 13, 2010
Date of Event
October 6, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY; (B)(4) THE FULL LEAD WAS RETURNED WHERE IT WAS DISCOVERED THAT THE DISTAL CONDUCTOR WAS DISTORTED. THE LEAD WAS STRETCHED. THE INNER INSULATION WAS KINKED/BUCKLED. THERE WAS BLOOD IN/ON THE HELIX MECHANISM. THERE WAS DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, THE STYLET GOT STUCK INSIDE THE RIGHT VENTRICULAR LEAD. THE LEAD WAS NOT USED, AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other