CAPSUREFIX NOVUS
MDR report key: 1923141
·
Received December 13, 2010
Report
- Report Number
- 2649622-2010-13265
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- October 6, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY; (B)(4) THE FULL LEAD WAS RETURNED WHERE IT WAS DISCOVERED THAT THE DISTAL CONDUCTOR WAS DISTORTED. THE LEAD WAS STRETCHED. THE INNER INSULATION WAS KINKED/BUCKLED. THERE WAS BLOOD IN/ON THE HELIX MECHANISM. THERE WAS DAMAGE AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THE STYLET GOT STUCK INSIDE THE RIGHT VENTRICULAR LEAD. THE LEAD WAS NOT USED, AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Other |