FDA Adverse Event Injury Summary report: N

SIGMA 300 DR

MDR report key: 1922971 · Received December 13, 2010

Report

Report Number
6000094-2010-02230
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO SYNCOPE, THAT RECENTLY THE PATIENT HAD THERAPY FOR BREAST CANCER, AND THAT THE PATIENT HAD NOT BEEN IN FOR FOLLOW UP FOR "SOME TIME". IT WAS ALSO REPORTED THAT TELEMETRY COULD NOT BE ESTABLISHED WITH THE DEVICE. THE DEVICE WAS REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 300 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SDR303 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R