3M ESPE ADPER SINGLE BOND PLUS ADHESIVE
Report
- Report Number
- 2110898-2010-00148
- Event Type
- Other
- Date Received
- December 9, 2010
- Date of Event
- January 1, 2010
- Report Date
- September 14, 2010
- Manufacturer
- 3M ESPE DENTAL PRODUCTS
- Product Code
- KLE
- PMA / PMN Number
- K962785
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
Narratives
METHODS, RESULTS AND CONCLUSIONS: THE THIRD DEVICE USED IN THIS EVENT WAS NOT RETURNED TO 3M ESPE AND NO LOT NUMBER WAS PROVIDED, THEREFORE, NO EVALUATION COULD BE CONDUCTED. THIS MANUFACTURER REPORT IS FOR DEVICE 3 OF 3 INVOLVED IN ONE SINGLE ADVERSE EVENT. REFER TO MANUFACTURER REPORT # 2110898-2010-00118 (DEVICE 1) AND #2110898-2010-00147 (DEVICE 2) FOR INFO ON THE OTHER TWO DEVICES. THE REPORTED EFFECTS ARE NOT EXPECTED FOR THE SUBJECT PRODUCTS; ALL HAVE A FAVORABLE CLINICAL USAGE HISTORY, WITH NO OTHER SIMILAR CASES REPORTED TO 3M ESPE. THE PLACEMENT AND CURING TECHNIQUES REPORTED BY THE DENTIST ARE NOT CONSISTENT WITH THE 3M ESPE INSTRUCTIONS FOR USE. IN ADDITION, THE LOT OF FILTEK SUPREME PLUS UNIVERSAL RESTORATIVE (DEVICE 2) USED IN THIS CASE WAS 1 YEAR PAST ITS STATED SHELF LIFE. IT IS UNK WHAT ROLE THE DENTIST'S TECHNIQUE AND/OR THE USE OF EXPIRED PRODUCT PLAYED IN THE CASE.
SEVERAL PTS (EXACT NUMBER AND INDIVIDUAL DETAILS NOT PROVIDED TO 3M ESPE) EXPERIENCED CRACKS IN THEIR TOOTH STRUCTURE (BOTH LINGUAL AND BUCCAL AREAS) AFTER DENTIST PERFORMED POSTERIOR RESTORATIONS WITH 3M ESPE FILTEK SUPREME PLUS FLOWABLE RESTORATIVE, 3M ESPE FILTEK SUPREME PLUS UNIVERSAL RESTORATIVE, AND 3M ESPE ADPER SINGLE BOND PLUS ADHESIVE. RESTORATIONS WERE EITHER NEW, OR WERE DONE TO REPLACE OLD AMALGAM FILLINGS. SOME PTS HAVE REQUIRED CROWNS OR ROOT CANALS; ALL PTS ARE REPORTED TO BE FINE. DENTIST INDICATED HER TECHNIQUE TO BE AS FOLLOWS: PLACES ABOUT 1MM OF FILTEK SUPREME PLUS FLOWABLE RESTORATIVE IN BOX AND LIGHT CURES; APPLIES 2 TO 3 COATS OF ADPER SINGLE BOND PLUS ADHESIVE AND LIGHTLY AIR DRIES BETWEEN EACH LAYER AND LIGHT CURES; PLACES FILTEK SUPREME PLUS UNIVERSAL RESTORATIVE GREATER OR LESS THAN 2MM IN INCREMENTAL DEPTH AND LIGHT CURES FOR 30 SECONDS. DENTIST ALSO INDICATED SHE USES A SLEEVE OVER THE LIGHT GUIDE AND TIP OF THE CURING LIGHT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 3M ESPE ADPER SINGLE BOND PLUS ADHESIVE | TOOTH BONDING, RESIN CEMENT, DENTAL | KLE | 3M ESPE DENTAL PRODUCTS | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |