FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 1922843 · Received December 13, 2010

Report

Report Number
2649622-2010-13104
Event Type
Injury
Date Received
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE PROXIMAL CONDUCTOR WAS FRACTURED. IT WAS NOTED THAT THE LEAD APPEARS TO HAVE BEEN IMPLANTED WITH A BEND IN IT AT THE FRACTURE SITE. IT WAS ALSO NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. NOISE WAS PRESENT ON THE RIGHT VENTRICULAR LEAD AND THERE WAS AN APPARENT LEAD FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. NOISE WAS PRESENT ON THE RIGHT VENTRICULAR LEAD AND OVERSENSING WAS ALSO REPORTED; THERE WAS AN APPARENT LEAD FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R 5594 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5594 IMPLANTABLE PACING LEAD