FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1922836 · Received December 13, 2010

Report

Report Number
2649622-2010-13101
Event Type
Injury
Date Received
December 13, 2010
Date of Event
October 1, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE PROXIMAL SEGMENT WAS RETURNED AND ANALYZED. THE DEFIBRILLATION CONDUCTOR WAS FRACTURED. THE DEFIBRILLATION CONDUCTOR WAS DISTORTED AND HAD FRACTURE (OVERSTRESS). THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY. THE INNER TUBING WAS KINKED/BUCKLED. THE OUTER TUBING OVERLAY WAS BREACHED CUT. THE OUTER INSULATION HAS COSMETIC DEPRESSION. THE EXPOSED DEFIBRILLATION COIL HAS WHITE SUBSTANCE ON IT.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD SHOWED HIGH IMPEDANCE AND APPARENT FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 48 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD