SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2010-13101
- Event Type
- Injury
- Date Received
- December 13, 2010
- Date of Event
- October 1, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE PROXIMAL SEGMENT WAS RETURNED AND ANALYZED. THE DEFIBRILLATION CONDUCTOR WAS FRACTURED. THE DEFIBRILLATION CONDUCTOR WAS DISTORTED AND HAD FRACTURE (OVERSTRESS). THERE WAS BLOOD/BODY FLUID ON THE OUTER TUBING OVERLAY. THE INNER TUBING WAS KINKED/BUCKLED. THE OUTER TUBING OVERLAY WAS BREACHED CUT. THE OUTER INSULATION HAS COSMETIC DEPRESSION. THE EXPOSED DEFIBRILLATION COIL HAS WHITE SUBSTANCE ON IT.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THE LEAD SHOWED HIGH IMPEDANCE AND APPARENT FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |