EON IPG
Report
- Report Number
- 1627487-2010-03254
- Event Type
- Malfunction
- Date Received
- November 12, 2010
- Date of Event
- October 15, 2010
- Report Date
- October 15, 2010
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION DIVISION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
THE PT RECEIVED HIS SCS SYSTEM ON (B)(6) 2007. IT WAS REPORTED THAT HIS PROGRAMMER IS FUNCTIONING AS DESIGNED; HOWEVER, HE IS UNABLE TO COMMUNICATE WITH HIS IPG VIA THE CHARGING SYSTEM. IN AN EFFORT TO RESOLVE THIS MATTER, A REPLACEMENT CHARGING SYSTEM WAS SHIPPED TO THE PT. FOLLOW-UP ON THE PT FOUND THAT THE PROBLEM PERSISTS WITH THE USE OF THE REPLACEMENT UNIT. A CONSULTATION HAS BEEN SCHEDULED FOR FURTHER INTERROGATION OF HIS SCS SYSTEM. THIS REPORT IS BEING SUBMITTED AS A PRECAUTIONARY MEASURE AS SIMILARLY REPORTED EVENTS HAVE LED TO THE EXPLANT OF THE IPG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EON IPG | TOTALLY IMPLANTABLE PULSE GENERATOR | LGW | ST JUDE MEDICAL - NEUROMODULATION DIVISION | 3716 | 65582 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |