SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2010-13070
- Event Type
- Malfunction
- Date Received
- December 13, 2010
- Date of Event
- September 20, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE FULL LEAD WAS RETURNED WHERE IT WAS DISCOVERED THAT THE EXPOSED DEFIBRILLATION COIL CONDUCTOR WAS SEVERELY DISTORTED WHICH IS SUSPECTED TO HAVE OCCURRED DURING THE IMPLANT PROCEDURE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE PHYSICIAN NOTED THAT THE 6947 LEAD COIL WAS SEPARATED AT THE PROXIMAL END OF THE RV COIL. UPON FURTHER INSPECTION BY THE PHYSICIAN IT WAS NOTED THAT THE SEPARATION WAS THROUGHOUT. A 6935 LEAD WAS PLACED INSTEAD AND PHYSICIAN FELT THE MOVEMENT WAS SNUG THROUGH THE INTRODUCER. THE 6935 WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE PHYSICIAN NOTED THAT THE 6947 LEAD COIL WAS SEPARATED AT THE PROXIMAL END OF THE RV COIL. UPON FURTHER INSPECTION BY THE PHYSICIAN IT WAS NOTED THAT THE SEPARATION WAS THROUGHOUT. A 6935 LEAD WAS PLACED INSTEAD AND PHYSICIAN FELT THE MOVEMENT WAS SNUG THROUGH THE INTRODUCER. THE 6935 WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 19 YR | Other |