FDA Adverse Event Injury Summary report: N

ADAPTA DR

MDR report key: 1922694 · Received December 13, 2010

Report

Report Number
6000144-2010-05973
Event Type
Injury
Date Received
December 13, 2010
Date of Event
September 1, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NVZ
PMA / PMN Number
P980035/S043
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES WERE FOUND. THERE WAS A LOOSE SET SCREW. (B)(4) NO ANOMALIES WERE FOUND. HOWEVER BLOOD WAS PRESENT ON THE HELIX AND SLEEVEHEAD. THE FULL LEAD WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD DISLODGED AND WAS UNABLE TO BE REATTACHED. THE ATRIAL LEAD WAS EXPLANTED AND REPLACED. THE SET SCREW ON EXISTING PACEMAKER WOULD NOT TIGHTEN ON NEW LEAD. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADAPTA DR IMPLANTABLE PULSE GENERATOR NVZ MEDTRONIC MED REL, INC. ADDR01 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD